Infant lung function: criteria for selecting tidal flow–volume loops

Karen Eline Stensby Bains, Hrefna Katrín Gudmundsdóttir, Martin Färdig, Erik Amnö, Christine M. Jonassen, Björn Nordlund, Eva Maria Rehbinder, Håvard O. Skjerven, Corina Silvia Rueegg, Riyas Vettukattil, Karin C. Lødrup Carlsen

Source: ERJ Open Res, 8 (4) 00165-2022; 10.1183/23120541.00165-2022
Journal Issue: October

Congress or journal article abstractFull text journal articlePDF journal article, handout or slides

Abstract

Background

Tidal flow–volume (TFV) loops are commonly recorded in infants during sleep, due to the more regular breathing patterns compared to the awake state. Standardised deselection of loops outside pre-specified ranges are based on periods of regular breathing, while criteria and available software for visual evaluation of TFV loops are lacking. We aimed to determine the reliability of standardised criteria for manual selection of infant TFV loops.

Methods

Using a pre-defined set of criteria, three independent raters manually evaluated TFV loops among 57 randomly selected awake healthy 3-month-old infants with available TFV measurements in the Scandinavian Preventing Atopic Dermatitis and ALLergies in children (PreventADALL) study. The TFV loops were sampled using the Eco Medics Exhalyzer D. Criteria for selecting TFV loops included reproducible shape and volume with only one peak in tidal expiratory flow (PTEF), excluding loops with no clear or uneven flow towards PTEF. By intraclass coefficient (ICC), the reliability of agreement between raters was determined for the time to PTEF (tPTEF) to expiratory time (tE) and other TFV loop parameters.

Results

Five infants had unsuccessful tests. Among the remaining 52 infants, the raters selected a median of 25, 26 and 15 loops per test. The ICCs (95% CI) were 0.97 (0.92–0.98) for tPTEF/tE, 0.99 (0.99–1.00) for respiratory rate, 0.98 (0.97–0.99) for tidal volume per kg and 0.98 (0.97–0.99) for expiratory volume, reflecting excellent agreement in all categories.

Conclusion

Manual TFV loop selection using standardised criteria provides a reliable alternative for lung function measures in awake infants with interrupted breathing cycles in a real-life setting.



Rating: 0
You must login to grade this presentation.

Share or cite this content

Citations should be made in the following way:
Karen Eline Stensby Bains, Hrefna Katrín Gudmundsdóttir, Martin Färdig, Erik Amnö, Christine M. Jonassen, Björn Nordlund, Eva Maria Rehbinder, Håvard O. Skjerven, Corina Silvia Rueegg, Riyas Vettukattil, Karin C. Lødrup Carlsen. Infant lung function: criteria for selecting tidal flow–volume loops. ERJ Open Res, 8 (4) 00165-2022; 10.1183/23120541.00165-2022

You must login to share this Presentation/Article on Twitter, Facebook, LinkedIn or by email.

Member's Comments

No comment yet.
You must Login to comment this presentation.