Performance of ventilators for noninvasive positive pressure ventilation in children

B. Fauroux, K. Leroux, G. Desmarais, D. Isabey, A. Clement, F. Lofaso, B. Louis (Paris, Puteaux, Creteil, Garches, France)

Source: Annual Congress 2008 - Progress and a new era for noninvasive mechanical ventilation
Session: Progress and a new era for noninvasive mechanical ventilation
Session type: Oral Presentation
Number: 361
Disease area: Paediatric lung diseases, Respiratory critical care

Congress or journal article abstractSlide presentation

Abstract

The aim of the study was to evaluate the performance characteristics of all the ventilators proposed for home noninvasive positive pressure ventilation in children in France.
Seventeen ventilators, (volume-targeted, n=1, pressure-targeted, n=12, and dual ventilators, n=4) were evaluated on a bench which simulated 6 different paediatric ventilatory patterns. For each ventilator, the quality of the inspiratory and expiratory trigger, and the ability to reach and maintain the preset pressures and volumes were evaluated with the 6 patient profiles.
The performance of the ventilators showed a great variability and depended on the type of trigger (flow or pressure), the type of circuit, and the patient profile. Differences between the preset and measured airway pressure and the tidal volume measured by the ventilator and on the bench were observed. Leaks were associated with the inability to detect the patient‘s inspiratory effort or autotriggering. No single ventilator was able to adequately ventilate the 6 paediatric profiles. Only few ventilators were able to adequately ventilate the profiles simulating the youngest patients.
A systematic paediatric bench evaluation is recommended for every ventilator proposed for home ventilation in order to detect any dysfunction and to guide the choice of the appropriate ventilator for a specific patient.


Rating: 0
You must login to grade this presentation.

Share or cite this content

Citations should be made in the following way:
B. Fauroux, K. Leroux, G. Desmarais, D. Isabey, A. Clement, F. Lofaso, B. Louis (Paris, Puteaux, Creteil, Garches, France). Performance of ventilators for noninvasive positive pressure ventilation in children. Eur Respir J 2008; 32: Suppl. 52, 361

You must login to share this Presentation/Article on Twitter, Facebook, LinkedIn or by email.

Member's Comments

No comment yet.
You must Login to comment this presentation.


Related content which might interest you:
Functional differences in bi-level pressure preset ventilators
Source: Eur Respir J 2001; 17: 268-273
Year: 2001



Monitoring during non-invasive positive pressure ventilation
Source: International Congress 2015 – PG3 Cardiorespiratory monitoring in critically ill patients
Year: 2015




Ventilators for noninvasive positive pressure ventilation: technical aspects
Source: ISSN=1025-448x, ISBN=1-904097-17-0, page=76
Year: 2001

Leak compensation in positive pressure ventilators: a lung model study
Source: Eur Respir J 2001; 17: 259-267
Year: 2001



NIV for acute respiratory failure: modes of ventilation and ventilators
Source: Eur Respir Mon; 2008: 41: 154–172
Year: 2008

Noninvasive positive pressure ventilation (NPPV) as a tool to help tracheal decanulation in children
Source: Annual Congress 2008 - Noninvasive ventilation in the chronic setting
Year: 2008


Home noninvasive positive pressure ventilation for severe COPD
Source: Virtual Congress 2020 – New technologies for the management of COPD and obstructive sleep apnea and the diagnosis of lung cancer
Year: 2020


Performance of ventilators for noninvasive positive-pressure ventilation in children
Source: Eur Respir J 2008; 31: 1300-1307
Year: 2008



Noninvasive positive pressure ventilation in burn patients
Source: Annual Congress 2011 - Noninvasive ventilation in the acute setting: education, organisation, H1N1, paediatrics, weaning, diagnostic procedures and special considerations
Year: 2011

Noninvasive positive pressure ventilation in COPD
Source: Breathe 2011; 8: 114-123
Year: 2011

Application of non-invasive positive pressure ventilation (NPPV) in a rural hospital
Source: Eur Respir J 2002; 20: Suppl. 38, 280s
Year: 2002

Domiciliary ventilators: from bench to bedside
Source: Eur Respir Mon; 2008: 41: 307–318
Year: 2008

Utility of noninvasive positive pressure ventilation (NPPV) for weaning COPD patients from invasive mechanical ventilation (IMV)
Source: Annual Congress 2009 - New trends in noninvasive ventilation for acute respiratory failure
Year: 2009

Longterm effects of noninvasive positive pressure ventilation (NPPV) in the obesity-hypoventilation-syndrome
Source: Eur Respir J 2004; 24: Suppl. 48, 546s
Year: 2004

Noninvasive ventilation: modes of ventilation
Source: ISSN=1025-448x, ISBN=1-904097-17-0, page=67
Year: 2001

Monitoring of transcutaneous PCO2 during initiation of noninvasive positive pressure ventilation (NPPV)
Source: Annual Congress 2007 - Physiological effects of noninvasive ventilation
Year: 2007



Bi-level positive airway pressure ventilation reduces the oxygen cost of breating in tracheostomized post-polio patients on invasive home ventilation
Source: Eur Respir J 2004; 24: Suppl. 48, 521s
Year: 2004

Macrocephaly and positive pressure ventilation in children
Source: Eur Respir J 2003; 22: Suppl. 45, 138s
Year: 2003

Analysis of ventilation and hemodynamic changes resulting from noninvasive bilevel pressure mechanical ventilation applied to newborn with acute pneumonia
Source: Annual Congress 2009 - Respiratory disorders of the newborn
Year: 2009

Continuous external negative pressure ventilation (CENPV) versus continuous positive pressure ventilation (CPPV) in intubated ARDS patients
Source: Eur Respir J 2002; 20: Suppl. 38, 200s
Year: 2002