Performance of ventilators for noninvasive positive pressure ventilation in children
B. Fauroux, K. Leroux, G. Desmarais, D. Isabey, A. Clement, F. Lofaso, B. Louis (Paris, Puteaux, Creteil, Garches, France)
Source: Annual Congress 2008 - Progress and a new era for noninvasive mechanical ventilation
Disease area: Paediatric lung diseases, Respiratory critical care
Abstract The aim of the study was to evaluate the performance characteristics of all the ventilators proposed for home noninvasive positive pressure ventilation in children in France. Seventeen ventilators, (volume-targeted, n=1, pressure-targeted, n=12, and dual ventilators, n=4) were evaluated on a bench which simulated 6 different paediatric ventilatory patterns. For each ventilator, the quality of the inspiratory and expiratory trigger, and the ability to reach and maintain the preset pressures and volumes were evaluated with the 6 patient profiles. The performance of the ventilators showed a great variability and depended on the type of trigger (flow or pressure), the type of circuit, and the patient profile. Differences between the preset and measured airway pressure and the tidal volume measured by the ventilator and on the bench were observed. Leaks were associated with the inability to detect the patient‘s inspiratory effort or autotriggering. No single ventilator was able to adequately ventilate the 6 paediatric profiles. Only few ventilators were able to adequately ventilate the profiles simulating the youngest patients. A systematic paediatric bench evaluation is recommended for every ventilator proposed for home ventilation in order to detect any dysfunction and to guide the choice of the appropriate ventilator for a specific patient.
Rating:
You must login to grade this presentation.
Share or cite this content
Citations should be made in the following way:
B. Fauroux, K. Leroux, G. Desmarais, D. Isabey, A. Clement, F. Lofaso, B. Louis (Paris, Puteaux, Creteil, Garches, France). Performance of ventilators for noninvasive positive pressure ventilation in children. Eur Respir J 2008; 32: Suppl. 52, 361
You must login to share this Presentation/Article on Twitter, Facebook, LinkedIn or by email.
Member's Comments
Related content which might interest you:
Related content which might interest you:
Functional differences in bi-level pressure preset ventilators Source: Eur Respir J 2001; 17: 268-273 Year: 2001
Monitoring during non-invasive positive pressure ventilation Source: International Congress 2015 – PG3 Cardiorespiratory monitoring in critically ill patients Year: 2015
Ventilators for noninvasive positive pressure ventilation: technical aspects Source: ISSN=1025-448x, ISBN=1-904097-17-0, page=76 Year: 2001
Leak compensation in positive pressure ventilators: a lung model study Source: Eur Respir J 2001; 17: 259-267 Year: 2001
NIV for acute respiratory failure: modes of ventilation and ventilators Source: Eur Respir Mon; 2008: 41: 154–172 Year: 2008
Noninvasive positive pressure ventilation (NPPV) as a tool to help tracheal decanulation in children Source: Annual Congress 2008 - Noninvasive ventilation in the chronic setting Year: 2008
Home noninvasive positive pressure ventilation for severe COPD Source: Virtual Congress 2020 – New technologies for the management of COPD and obstructive sleep apnea and the diagnosis of lung cancer Year: 2020
Performance of ventilators for noninvasive positive-pressure ventilation in children Source: Eur Respir J 2008; 31: 1300-1307 Year: 2008
Noninvasive positive pressure ventilation in burn patients Source: Annual Congress 2011 - Noninvasive ventilation in the acute setting: education, organisation, H1N1, paediatrics, weaning, diagnostic procedures and special considerations Year: 2011
Noninvasive positive pressure ventilation in COPD Source: Breathe 2011; 8: 114-123 Year: 2011
Application of non-invasive positive pressure ventilation (NPPV) in a rural hospital Source: Eur Respir J 2002; 20: Suppl. 38, 280s Year: 2002
Domiciliary ventilators: from bench to bedside Source: Eur Respir Mon; 2008: 41: 307–318 Year: 2008
Utility of noninvasive positive pressure ventilation (NPPV) for weaning COPD patients from invasive mechanical ventilation (IMV) Source: Annual Congress 2009 - New trends in noninvasive ventilation for acute respiratory failure Year: 2009
Longterm effects of noninvasive positive pressure ventilation (NPPV) in the obesity-hypoventilation-syndrome Source: Eur Respir J 2004; 24: Suppl. 48, 546s Year: 2004
Noninvasive ventilation: modes of ventilation Source: ISSN=1025-448x, ISBN=1-904097-17-0, page=67 Year: 2001
Monitoring of transcutaneous PCO2 during initiation of noninvasive positive pressure ventilation (NPPV) Source: Annual Congress 2007 - Physiological effects of noninvasive ventilation Year: 2007
Bi-level positive airway pressure ventilation reduces the oxygen cost of breating in tracheostomized post-polio patients on invasive home ventilation Source: Eur Respir J 2004; 24: Suppl. 48, 521s Year: 2004
Macrocephaly and positive pressure ventilation in children Source: Eur Respir J 2003; 22: Suppl. 45, 138s Year: 2003
Analysis of ventilation and hemodynamic changes resulting from noninvasive bilevel pressure mechanical ventilation applied to newborn with acute pneumonia Source: Annual Congress 2009 - Respiratory disorders of the newborn Year: 2009
Continuous external negative pressure ventilation (CENPV) versus continuous positive pressure ventilation (CPPV) in intubated ARDS patients Source: Eur Respir J 2002; 20: Suppl. 38, 200s Year: 2002