Validation of a multiple breath inert gas washout technique using a lung phantom

S. Gonem, F. Singer, S. Siddiqui, P. Gustafsson (Leicester, United Kingdom; Bern, Switzerland; Skövde, Sweden)

Source: Annual Congress 2012 - Various issues in clinical physiology
Session: Various issues in clinical physiology
Session type: Thematic Poster Session
Number: 3798
Disease area: Airway diseases, Respiratory critical care

Congress or journal article abstract

Abstract

Background: Multiple breath inert gas washout (MBW) is a technique for assessing ventilation heterogeneity by measuring the efficiency with which an inert tracer gas is washed out of the lungs.
Aims and objectives: We aimed to use a lung phantom to validate a previously published method for MBW (Horsley AR, et al. Thorax. 2008; 63(2): 135-140).
Methods: We built a one-compartment lung model consisting of an enclosed clear acrylic glass tank, partly filled with water at 37oC. The tank was fitted with an off-centre vertical partition that divided it into a larger and a smaller section, but did not reach the base of the tank, thus allowing the two sides to communicate. The lid of the larger section was connected to a bi-level positive airway pressure ventilator which exerted alternating high and low pressures on the water surface, thus causing the water level in the smaller section to alternately rise and fall, simulating diaphragmatic movement. FRC, tidal volume (Vt) and respiratory rate (RR) were adjusted by altering the volume of water within the phantom lung and the ventilator settings. We performed MBW on the phantom lung at a variety of values of FRC, Vt and RR. Each experiment was performed in triplicate. Measured FRC was then compared to the true FRC in each case.
Results: Measured FRC was within 5% of true FRC in all but one experiment. The intraclass correlation coefficient for the triplicate measurements was 0.999, suggesting that the MBW technique is highly repeatable.
Conclusion: We conclude that MBW performed using an Innocor gas analyser is highly repeatable and accurate. Simple one-compartment lung phantoms appear to be a promising method for the validation of MBW systems.


Rating: 0
You must login to grade this presentation.

Share or cite this content

Citations should be made in the following way:
S. Gonem, F. Singer, S. Siddiqui, P. Gustafsson (Leicester, United Kingdom; Bern, Switzerland; Skövde, Sweden). Validation of a multiple breath inert gas washout technique using a lung phantom. Eur Respir J 2012; 40: Suppl. 56, 3798

You must login to share this Presentation/Article on Twitter, Facebook, LinkedIn or by email.

Member's Comments

No comment yet.
You must Login to comment this presentation.


Related content which might interest you:
Validation of a modified photo-acoustic gas analyser for in-vitro multiple breath washout in young children
Source: Annual Congress 2010 - Lung function from infancy to adulthood
Year: 2010


Modifications to a photoacoustic gas analyser increases accuracy of functional residual capacity measurements by multiple breath washout of SF6 in an infant lung model
Source: Annual Congress 2008 - Clinical applications of ventilatory function tests in paediatric pulmonology
Year: 2008


Comparison of measures of ventilation heterogeneity derived from multiple breath inert gas washout
Source: Annual Congress 2012 - Pathophysiological mechanisms in disease: new insights
Year: 2012

In vitro validation of nitrogen multiple breath washout using ultrasonic equipment
Source: Annual Congress 2011 - Advances in lung function testing from infancy to adulthood
Year: 2011

Comparison of helium and sf6 multiple breath gas washout techniques in spontaneously breathing infants
Source: Eur Respir J 2003; 22: Suppl. 45, 383s
Year: 2003

Effects of using a mask vs mouthpiece on the multiple breath inert gas washout technique
Source: International Congress 2015 – Lung function: waiting to exhale...
Year: 2015

Routine measurement of the LCI in CF with an ultrasonic device for multiple breath nitrogen washout
Source: Annual Congress 2012 - New understanding of childhood lung disease through physiological measurement
Year: 2012

Closed circuit rebreathing to achieve inert gas wash-in for multiple breath wash-out
Source: ERJ Open Res 2016
Year: 2016



Simultaneous measurement of multiple breath washout and functional lung MRI
Source: Virtual Congress 2021 – New insights into lung function testing
Year: 2021

Double-tracer gas single breath washout - Variability and reproducibility compared with classical nitrogen washout tests
Source: Annual Congress 2013 –New issues in lung function testing
Year: 2013


Comparison of a new nitrogen multiple breath washout method to mass spectrometer SF6 washout in cystic fibrosis subjects
Source: Annual Congress 2011 - Advances in lung function testing from infancy to adulthood
Year: 2011


Interpretation of multiple breath washout (MBW) measurements of lung function using mathematical modelling and hyperpolarised 3He gas MRI
Source: Virtual Congress 2020 – Advances in cystic fibrosis: from microbiome to MRI
Year: 2020




Validation of a new non invasive breath by breath method for the quantification of autoPEEP
Source: Eur Respir J 2006; 28: Suppl. 50, 817s
Year: 2006

Simultaneous measurement of inhaled air and exhaled breath by double multicapillary column ion-mobility spectrometry, a new method for breath analysis: results of a feasibility study
Source: ERJ Open Res, 8 (1) 00493-2021; 10.1183/23120541.00493-2021
Year: 2022



Equipment for multiple breath washout
Source: Eur Respir J 2010; 35: 458
Year: 2010


Automated detection of the alveolar plateau during multiple breath nitrogen washout
Source: Annual Congress 2010 - Advances in lung function II
Year: 2010

The impact of gas analysis technique on the measurement of single breath gas transfer (TLCO)
Source: International Congress 2015 – Lung function: effects on method, machine and disease on the measurement
Year: 2015


Influence of gas concentration and measurement interval on the reproducibility of non-invasive cardiac output determination by inert gas rebreathing in pulmonary healthy patients
Source: Annual Congress 2011 - Innovative methods in exercise testing
Year: 2011

Multiple breath washout techniques
Source: ERS Skills Course 2016
Year: 2016